If you are viewing on a smaller screen or with screen zoom, please use the horizontal scroll bar at the bottom of this table to see all information.
SERIES | 100 | |||||||||
PRESSURE RANGES |
30Vac | -30 inHg to 0 psig | 5 | 0 psig to 5 psig | 200 | 0 psig to 200 psig | 3000 | 0 psig to 3000 psig | 15A | 0 psia to 15 psia |
30/15 | -30 inHg to 15 psig | 10 | 0 psig to 10 psig | 300 | 0 psig to 300 psig | 4000 | 0 psig to 4000 psig | 30A | 0 psia to 30 psia | |
30/30 | -30 inHg to 30 psig | 15 | 0 psig to 15 psig | 500 | 0 psig to 500 psig | 5000 | 0 psig to 5000 psig | 60A | 0 psia to 60 psia | |
30/45 | -30 inHg to 45 psig | 25 | 0 psig to 25 psig | 600 | 0 psig to 600 psig | 6000 | 0 psig to 6000 psig | 100A | 0 psia to 100 psia | |
30/100 | -30 inHg to 100 psig | 30 | 0 psig to 30 psig | 750 | 0 psig to 750 psig | 7500 | 0 psig to 7500 psig | 150A | 0 psia to 150 psia | |
30/150 | -30 inHg to 150 psig | 60 | 0 psig to 60 psig | 1000 | 0 psig to 1000 psig | 10000 | 0 psig to 10000 psig | 200A | 0 psia to 200 psia | |
30/200 | -30 inHg to 200 psig | 100 | 0 psig to 100 psig | 1500 | 0 psig to 1500 psig | 15000 | 0 psig to 15000 psig | 300A | 0 psia to 300 psia | |
30/300 | -30 inHg to 300 psig | 150 | 0 psig to 150 psig | 2000 | 0 psig to 2000 psig | |||||
psig = gauge pressure psia = absolute pressure Other ranges available on special request | ||||||||||
ACCURACIES | 1 | ±0.5 % full scale (BFSL) | 2 | ±0.25 % full scale (BFSL) | ||||||
OUTPUT SIGNAL | 1 | 4 mA to 20 mA, 2-wire | ||||||||
PROCESS CONNECTIONS |
1 | 1/8" NPT Male | 3 | SAE J1926-3:7/16-20 Adjustable | 9 | SAE J1926-1:7/16-20 | ||||
2 | 1/4" NPT | 4 | 1/8" NPT Female | 10 | G1/4 Male | |||||
ELECTRICAL CONNECTIONS |
1 | 36" cable (connected to option 7) | 6 | 1/2" NPT conduit (with 36" cable) | 25 | M12 x 1 (4-pin) | ||||
2 | 4-pin Bendix | 7 | Mini-Hirschmann (DIN EN 175301-803 Form C) | 36 | Integral cable 36" | |||||
3 | 6-pin Bendix | |||||||||
OPTION | ST8 | Threaded orifice |
If you are viewing on a smaller screen or with screen zoom, please use the horizontal scroll bar at the bottom of this table to see all information.
SPECIFICATIONS | |
Output signal | 4 mA to 20 mA, 2-wire |
Pressure ranges | Vacuum through 0 psig to 15,000 psig Absolute from 0 psia to 15 psia through 0 psia to 300 psia |
Accuracy | ±0.5% full scale (BFSL); optional ±0.25% full scale (BFSL); (Includes the effects of non-linearity, hysteresis, non-repeatability, zero point and full scale errors) |
Stability | ≤ ±0.2% full scale for 1 year, non-accumulating |
Adjustment | ≤ ±10% full scale for zero and span |
Response time | ≤ 1 ms (between 10% and 90% full scale) |
Pressure cycle limit | 150 Hz |
Service life | > 100,000,000 load cycles |
Temperature ranges | Compensated 32 °F to 176 °F (0 °C to 80 °C) Effect ±0.017% full scale/ °F for zero and span Media -22 °F to 212 °F (-30 °C to 100 °C) Ambient -40 °F to 185 °F (-40 °C to 85 °C) Storage -40 °F to 212 °F (-40 °C to 100 °C) |
Power requirement* | 10 Vdc to 30 Vdc (4 mA to 20 mA, 2-wire) |
Load limitations | ≤ (Vpower supply -10)/.020 Amp |
Proof pressure | 3 times full scale for ranges 0 psi to 5 psi through 0 psi to 200 psi 1.75 times full scale for ranges 0 psi to 300 psi through 0 psi to 10,000 psi 1.5 times full scale for 0 to 15,000 psi |
Burst pressure | 3.8 times full scale for ranges 0 psi to 5 psi through 0 psi to 200 psi 4 times full scale for ranges 0 psi to 300 psi through 0 psi to 10,000 psi 3 times full scale for 0 psi to 15,000 psi |
Measuring element | 316 stainless steel for vacuum through 300 psi; 17-4PH stainless steel for ≥500 psi |
Connection | 316 stainless steel |
Housing material | 316 stainless steel |
Environmental rating | IP65 |
Electromagnetic rating | CE compliant to EMC norm EN 61326:2014/A1:1998 RFI, EMI and ESD protection |
Electrical protection | Reverse polarity, over-voltage and short circuit protection |
Shock | 1000 g’s according to IEC 60068-2-27 |
Vibration | 20 g’s according to IEC 60068-2-6 |
Weight | Approximately 3.5 oz. |
NOSHOK offers a wide variety of standard & elevated pressure, and reduced pressure replaceable and non-replaceable diaphragm seals, designed to isolate and/or protect the pressure measuring instrument from corrosive, viscous, contaminated or very high temperature process media.
View NOSHOK's Diaphragm Seal offering
NOSHOK Digital Indicators provide local indication for pressure, level and temperature transmitters.
View NOSHOK's Digital Indicators offering
NOSHOK's Distribution Manifolds connect multiple instruments to one diaphragm seal, eliminating multiple connections that lead to possible leak paths.
View NOSHOK Distribution Manifolds
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
Steam Syphons Temperature 1/4" NPT vs. Pressure charts
Steam Syphons Temperature 1/2" NPT vs. Pressure charts
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov