Use the PT Series Part Number Crossover Tool to convert from 100, 200, 300 and 615 Series transmitters to the new PT Series transmitters.
If you are viewing on a smaller screen or with screen zoom, please use the horizontal scroll bar at the bottom of this table to see all information.
GAUGE SIZES | 25 | 2-1/2" | 40 | 4" | ||
CASE TYPES | 200 | Steel, bottom connection (304SS for 4") | 224 | ABS, panel mount | ||
210 | Steel, back connection | 234 | Gas pressure test kit ** | |||
PRESSURE RANGES |
15 inH2O Vac | -15 inH2O to 0 inH2O | 200 inH2O | 0 inH2O to 200 inH2O | 3psi | 0 psi to 3 psi |
30 inH2O Vac | -30 inH2O to 0 inH2O | 10 oz/in2 | 0 oz/in2 to 10 oz/in2 | 5 psi | 0 psi to 5 psi | |
60 inH2O Vac | -60 inH2O to 0 inH2O | 15 oz/in2 | 0 oz/in2 to 15 oz/in2 | 10 psi | 0 psi to 10 psi | |
100 inH2O Vac | -100 inH2O to 0 inH2O | 30 oz/in2 | 0 oz/in2 to 30 oz/in2 | 40 mbar | 0 mbar to 40 mbar | |
10 inH2O | 0 inH2O to 10 inH2O | 35 oz/in2 | 0 oz/in2 to 35 oz/in2 | 60 mbar | 0 mbar to 60 mbar | |
15 inH2O | 0 inH2O to 15 inH2O | 60 oz/in2 | 0 oz/in2 to 60 oz/in2 | 160 mbar | 0 mbar to 160 mbar | |
30 inH2O | 0 inH2O to 30 inH2O | 100 oz/in2 | 0 oz/in2 to 100 oz/in2 | 250 mbar | 0 mbar to 250 mbar | |
60 inH2O | 0 inH2O to 60 inH2O | 160 oz/in2 | 0 oz/in2 to 160 oz/in2 | 400 mbar | 0 mbar to 400 mbar | |
100 inH2O | 0 inH2O to 100 inH2O | 20 oz/35 inH2O | 0 oz/in2/inH2O to 20 oz/in2 inH2O | 600 mbar | 0 mbar to 600 mbar | |
160 inH2O | 0 inH2O to 160 inH2O | 32 oz/55 inH2O | 0 oz/in2/inH2O to 32 oz/in2 inH2O | |||
CONNECTION SIZE | 1/4 | 1/4" NPT | ||||
OPTIONS | BLRF | Black Rear Flange | PL | Acrylic Lens | SSCR | Stainless Steel Cover Ring |
SSRF | 304SS Rear Flange | RL | Recalibrator Lens | CCR | Chrome Cover Ring | |
BLFF | Black Front Flange *** | SP | Red Set Pointer | LM | Laser Marking | |
SSFF | 304SS Front Flange | MIP | Maximum Indicatinrg Pointer | ST | Stainless Steel Tagging | |
CFF | Chrome Front Flange *** | OP | Over Pressure Protection | BP3 | Brass Press Fit Orifice 0.3 mm | |
SSC | Stainless Steel Case | SSBU | Stainless Steel Bezel & U-Clamp | BT3 | Brass Threaded Orifice 0.3 mm | |
GL | Glass Lens * | BBU | Black Beze & U-Clamp | |||
SG | Safety Glass Lens * | BCR | Black Cover Ring |
Note: O2 cleaning is not available for 200 Series gauges.
* A Steel, Stainless or chrome cover ring must be additionally ordered when lenses other than acrylic are utilized on all 200 Series
** Only available in 2-1/2" size, 20 oz/35 inH2O
*** Not available on 25-200 model
Please consult your local NOSHOK Distributor or NOSHOK, Inc. for availability and delivery information.
NOTES: See Standard Connection Sizes
Shaded portions of this part number example are not required unless a non-standard connection size or an option is needed.
If you are viewing on a smaller screen or with screen zoom, please use the horizontal scroll bar at the bottom of this table to see all information.
SERIES | SPECIFICATIONS | |
Pressure ranges | 200 Series (all) | Extreme low pressure vacuum ranges through 0 psi to 10 psi |
Accuracy | 25-200, 25-210, 25-224 | ±1.6% full scale |
25-206, 25-216 | ±2.5% full scale | |
40-200 | ±1.6% full scale; optional ±1% full scale | |
Temperature ranges* | 200 Series (all) | Media -4 °F to 176 °F (-20 °C to 80 °C) Ambient -4 °F to 140 °F (-20 °C to 60 °C) |
Measuring element | 25-200, 25-210, 25-224, 40-200 | Copper alloy diaphragm capsule |
Connection | 200 Series (all) | 1/4″ NPT, brass |
Case | 25-200, 25-210 | Black painted steel |
25-206, 25-216, 25-224 | Black ABS (Acryl Nitril Butadien Styrol) with 25-224 includes zinc-plated steel panel mount clamp | |
40-200 | 304 Stainless steel | |
Bezel | 40-200 | 304 Stainless steel |
Lens | 25-200, 25-206, 25-210, 25-216, 25-224 | Acrylic |
40-200 | Instrument glass | |
Pointer | 200 Series (all) | Black finished aluminum |
Dial | 200 Series (all) | Aluminum, white background with black scale. UV resistant. |
Movement | 25-200, 25-210, 25-224, 40-200 | Brass and nickel-silver with highly polished bearing surfaces |
25-206, 25-216 | Cu-Alloy |
Working Pressure Limitations
Dynamic Pressure
The working pressure should be limited to 60% of the dial range.
Static Pressure
The working pressure, when no sharp fluctuations occur, should be limited to 90% of the dial range.
AVAILABLE
CERTIFICATES
* A steel or stainless cover ring must be additionally ordered when lenses other than plexiglass are utilized on all 200 Series models
WARNING:This product can expose you to chemicals including Lead and Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
AVAILABLE
CERTIFICATES
* A steel or stainless cover ring must be additionally ordered when lenses other than plexiglass are utilized on all 200 Series models
WARNING:This product can expose you to chemicals including Lead and Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
AVAILABLE
CERTIFICATES
WARNING:This product can expose you to chemicals including Lead and Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
AVAILABLE
CONSULT FACTORY
CERTIFICATES
* A steel or stainless cover ring must be additionally ordered when lenses other than plexiglass are utilized on all 200 Series models.
WARNING:This product can expose you to chemicals including Lead and Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
NOSHOK offers a wide variety of standard & elevated pressure, and reduced pressure replaceable and non-replaceable diaphragm seals, designed to isolate and/or protect the pressure measuring instrument from corrosive, viscous, contaminated or very high temperature process media.
View NOSHOK's Diaphragm Seal offering
NOSHOK's Distribution Manifolds connect multiple instruments to one diaphragm seal, eliminating multiple connections that lead to possible leak paths.
View NOSHOK Distribution Manifolds
Please consult the factory for availability.
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
Steam Syphons Temperature 1/4" NPT vs. Pressure charts
Steam Syphons Temperature 1/2" NPT vs. Pressure charts
WARNING: This product can expose you to chemicals including Lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING: This product can expose you to chemicals including Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov
WARNING:This product can expose you to chemicals including Lead and Nickel, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov